Alloplastic temporomandibular joint replacement—what does the future hold?
Review Article

Alloplastic temporomandibular joint replacement—what does the future hold?

Louis G. Mercuri1,2,3

1Department of Orthopaedic Surgery, Rush University Medical Center, Chicago, IL, USA; 2TMJ Concepts, Ventura, CA, USA; 3Department of Bioengineering, University of Illinois Chicago, Chicago, IL, USA

Correspondence to: Louis G. Mercuri, DDS, MS. 604 Bonnie Brae Place, River Forest, IL 60305, USA. Email: lgm@tmjconcepts.com.

Abstract: Salvage management of end-stage TMJ pathologic conditions are considered indications for alloplastic temporomandibular joint replacement (TMJR). The primary goal of TMJR is the safe and effective restoration of mandibular function and form. In order to provide successful function, form, as well as long-term subjective and objective outcomes, any present or future TMJR device must be able to safely and effectively manage the anatomic, functional, and aesthetic discrepancies presented to it for reconstruction. To accomplish this, all TMJR devices, present and future, must demonstrate that they are constructed using biologically compatible materials, designed and manufactured to bear the loads imposed on their components by function, and that they are accurately analyzed biomechanically and clinically tested to assure long-term efficacy for patients. In the future, biomechanical laboratory and clinical outcome studies should be a guide not only to surgeons and patients, but also regulatory, health care and the indemnification communities of interest. Additive manufacturing (3D printing), augmented reality (AR), artificial intelligence (AI) and robot-assisted surgery (RAS) will become important tools for the general growth and development of the specialty of oral and maxillofacial surgery. What role these will play in the future of TMJ disorder diagnosis, non-surgical and surgical management remains to be determined. This paper will discuss the past and present iterations of TMJR devices, the embodiments, the successes, and failures as a guide to what the future may hold for researchers and clinicians with regard to the development of the next generation of TMJR devices.

Keywords: Temporomandibular joint replacement (TMJR); alloplastic TMJ reconstruction; end-stage TMJ pathology


Received: 08 June 2020; Accepted: 22 July 2020; Published: 04 September 2020.

doi: 10.21037/fomm-2020-tjddm-02


“If you always do what you always did, you’ll always get what you’ve always got.”—Henry Ford

“Continuous improvement is better than delayed perfection.”—Mark Twain

“We cannot become what we want to be by remaining what we are.”—Max DePree

“Perfection is not attainable. But if we chase perfection, we can catch excellence.”—Vince Lombardi

For years, autogenous costochondral bone was considered the “gold standard” for temporomandibular joint (TMJ) reconstruction. Commercial or “stock” alloplastic temporomandibular joint replacement (TMJR) devices were restricted to the management of ankylosis, reconstruction after ablative surgery, trauma, or end-stage joint disease cases. However, the anatomical architectural aberrations and local pathologic reactions created by the failure of materials like Proplast-Teflon and silicone rubber TMJ implants led to the development of patient fitted or “custom” TMJR devices in 1990 (1).

Since that time, the US Food and Drug Administration (FDA) has approved both custom and improved stock TMJR systems that employ orthopedic alloplastic joint replacement biomechanical concepts and material embodiments that have proven over decades to be not only safe, but also effective in the management of end-stage joint disease (2). However, future advancements in material science, design, manufacturing as well as surgical techniques and equipment will foster improvements in all joint replacement devices.

Based on evidence from material science, orthopedic and TMJR literature, this paper will discuss TMJR relative to future demand for these devices, the training of surgeons to meet those demands, the materials, design/manufacturing processes, biomechanical analysis and clinical testing, regulatory requirements and clinical outcomes.

The demand for total alloplastic joint replacement has increased over the years in both orthopedic and maxillofacial surgery. This is the result of improved designs/materials, reported enhanced successful subjective and objective outcomes over autogenous tissue grafts, and most importantly reports of increased quality of life for patients receiving these devices.

Onoriobe et al. reported a 38% increase in TMJR cases between 2005 and 2014 (3). Using the same statistical metrics as in the Kurtz et al. total hip and knee utilization projection study, these authors projected a 58% increased demand for TMJR by 2030 (4).

In a survey of US oral and maxillofacial training program directors, Lotesto et al. reported an average of 0 to 6 TMJR cases were performed annually at those programs. Further, the respondents significantly disagreed when asked if residents nationwide will be competent to operated and manage TMJR patients. Therefore, 46.9% thought that increased TMJR training was necessary to improve future outcomes (5).

In order to assess the confidence level of US trainees in the performance of arthrocentesis and arthroscopy and TMJR upon completion of residency, Momin et al. conducted a follow-up to the Lotesto survey. The results revealed that the confidence level of responders correlated directly with their exposure to and experience with a TMJ related procedure during residency (6).

Therefore, in the future there should be concern that the next generation of oral and maxillofacial surgeons may not be sufficiently trained to manage the projected TMJR demand unless there is modification in the training requirements and experience in the US and potentially world-wide.

The general considerations for the development of a biomedical device were described by Lantada and Morgado as the existence of a relevant medical need, the anatomical and biological regard for the selection of biomaterials with appropriate mechanical and chemical properties, the availability of appropriate sterilization methods, and the knowledge of the consequences of wear and corrosion (Table 1) (7) .

Table 1

General considerations for the development of a biomedical device

1. Existence of a relevant medical need
2. Anatomical and biological regard for the selection of biomaterial with appropriate mechanical and chemical properties
3. Availability of sterilization methods
4. Knowledge of the consequences of wear and corrosion

Lantada AD, Morgado PL. General Considerations for the Development of Biomedical Devices. In: Lantada AD (ed). Handbook on Advanced Design and Manufacturing Technologies for Biomedical Devices. New York: Springer, 2013;19-45.

With regard to the general requirements for joint related devices, Katti stated that they should be composed of biocompatible materials that do not produce inflammation or systemic toxicity beyond acceptable levels, are designed to provide the mechanical properties for their intended function, and are able to be fabricated cost-effectively, employing rigorously tested and vetted processing and manufacturing procedures (Table 2) (8).

Table 2

General requirements for joint replacement devices

1. Utilization of biocompatible materials that do not produce inflammation or systemic toxicity beyond acceptable levels
2. Designed to provide the mechanical properties for their intended function
3. Able to be fabricated cost-effectively employing rigorously tested and vetted processing and manufacturing procedures

Katti K. Biomaterials in total joint replacement. Colloids and surfaces. Biointerfaces 2004;39:133-42.

Specifically for TMJR, Mercuri described the criteria for a successful TMJR device as, it be designed and manufactured from material biocompatible with the implant environment; it be able to withstand repetitive loading and unloading delivered over the full range of TMJ function while remaining stable in situ; and that the surgery to implant the TMJR be performed aseptically and for the proper indications (Table 3) (9).

Table 3

Criteria for a successful TMJR device

1. Designed and manufactured from material biocompatible with the implant environment
2. Able to withstand repetitive loading and unloading over the full range of TMJ function
3. Components must be stable in situ
4. Surgery must be performed for the appropriate indications and aseptically

Mercuri LG. The Use of Alloplastic Prostheses for Temporomandibular Joint Reconstruction. J Oral Maxillofac Surg 2000;58:70-5. TMJR, temporomandibular joint replacement.

Elledge et al. published a review of emerging TMJR systems (10). The authors reported that at the time of that publication, 15 countries had already created or were in the process of developing 27 TMJR systems, 21 of which were custom designed and manufactured. However, only 4 had been sanctioned by a regulatory agency. Twenty-one devices employed an ultra-high-molecular-weight polyethylene (UHMWPE) fossa-bearing surface. Ten contained all-titanium alloy condyles. Nineteen manufacturers report that a titanium alloy was used for the ramus portion of the ramus/condyle component. Twelve reported preclinical biomechanical testing, but no clinical outcomes were reported for 9 of the 27 systems. The two US FDA-approved systems (TMJ Concepts, Ventura, CA (11) and Zimmer Biomet, Jacksonville, FL (12) had been providing the market share of TMJR devices worldwide. However, tariffs, agent, and distributor fees, as well as the economic conditions and healthcare conventions in many countries contributed to an increased premium for utilizing US devices. These factors, along with the lure of 3D printing, appeared to have led clinicians, researchers, and device manufacturers in many countries to start developing what were considered “less costly” TMJR devices.

After reviewing the data from this study, Elledge et al. discovered that all TMJR systems uncovered were not equivalent with regards to design, materials, manufacturing practices, biomechanical testing, clinical outcomes reporting or regulatory status (10).

Considering TMJR device design, the uniqueness of the mandibular ramus and temporal glenoid fossa anatomies do not lend themselves to the use of stock modular replacement components. The bony anatomy of the pelvis, femur, and tibia lend themselves to the use of modular stock components. Therefore, for the foreseeable future, all TMJR components must be designed to be fixated and secured for stabilization to the maxillofacial skeletal host ramus and temporal bones with bicortical screws (13).

Biomechanically, finite element analysis confirmed that the maximum functional forces placed on a TMJR ramus/condyle component are concentrated at the most superior screw hole (14). Further, Hsu et al. (15) and Ramos et al. (16) reported studies demonstrating the importance of the screw fixation along the full length of lateral ramus to maintain stable fixation of the ramus condyle component. Therefore, any future ramus/condyle component design that does not require the use of the most superior fixation screw or does not utilize the full length of the vertical mandibular ramus for screw fixation, should not be considered biomechanically sound.

Additive manufactured or 3D printed devices are being integrated into surgical practice. Applications vary from development of anatomical models for surgical planning to surgical guides as well as implant devices themselves. As of the date of their publication, Elledge et al. reported that 7 TMJR systems involved the use of 3D printed metal components. Three were fashioned using direct metal laser sintering (DMLS), 1 by electron beam melting (EBM), and 3 by selective laser melting (SLM) (10).

In a systematic review of the 3D printing process from selected papers in the medical literature, Tack et al. found that the major clinical advantages were reported to be reduced surgical times, improved medical outcomes, and decreased radiation exposure. However, they concluded that there was lack of data supporting most of those advantages (17). The major disadvantage reported was increased cost, making it questionable whether 3D printing is cost effective for all patients and applications (18,19).

Biomechanical concerns have arisen associated with the use of additive 3D printing of metallic medical devices. These include, porosity—the size and number of pores developed in 3D printed component can result in a low-density material making it more prone to fracture under functional loading; density control—when a component undergoes functional cyclical stress, its density will determine whether the part will crack and fail when loaded; residual stress control—temperature changes during the 3D printing process can lead to residual stress resulting in deformation of the component; cracking and warping—this occur when the melted metal cools down after printing. Cooling causes contraction which makes edges of the component deform and the part cracking under stress. Cracking may also occur if the powder material was not properly melted; and post processing surface roughness—3D printed components are often developed with rough surfaces. In order to achieve the desired final finish further machining, grinding, or polishing is required. Those operations can further damage the surface of the component by exposing deeper, larger pores which ultimately may provide harbor for contaminant organisms (Table 4) (20).

Table 4

Biomechanical concerns with the use of 3D printing of metallic medical devices

1. Porosity control
2. Density control
3. Residual stress control
4. Cracking and Warping
5. Post processing surface roughness

Finally, the biggest future challenges facing the 3D printing industry have been reported to be equipment costs, limited material availability, manufacturing resources and costs, lack of in-house additive manufacturing expertise, limited accuracy from case to case, longer production timelines, liability implications, and lack of formal industrial standards (Table 5) (21).

Table 5

Future challenges facing the 3D printing industry

1. Equipment costs
2. Limited materials availability
3. Manufacturing resources and costs
4. Lack of in-house additive manufacturing expertise
5. Limited accuracy from case to case
6. Longer production timelines
7. Liability implications
8. Lack of formal standards

Tack et al. concluded that further research is required to determine whether the increased intervention costs associated with 3D printing can be balanced by observable advantages (17). Therefore, TMJR manufacturers planning to use this technology in the future should perform cost–benefit analyses.

When deciding on the materials to be used in the manufacture of future TMJR devices, researchers and manufactures must utilize materials that have the most beneficial physical and biocompatible characteristics.

At present, ultra-high molecular weight polyethylene (UHMWPE) remains the “gold standard” in orthopedic joint replacement devices as the bearing surface for the stable component of hip and knee replacement devices (i.e., acetabular cup, tibial plateau) (13) UHMWPE combines a low coefficient of friction with outstanding impact strength. The property of creep (cold flow), rather than particulation, making this self-lubricating polymer an excellent choice for a bearing surface of any joint replacement device.

Biofilm infection is the most common cause of alloplastic joint replacement failure in both TMJR and orthopedic joint replacement. The material qualities that make UHMWPE the ideal bearing surface also make it the most likely target for biofilm formation. Biofilms form with the attachment of planktonic microorganisms to a surface. The first colonist bacteria of a biofilm initially adhere to UHMWPE surfaces by weak van der Waals forces and hydrophobic effects (22-25).

Chemical modification has become the main strategy for biofilm prevention on indwelling medical devices. Antibiotics, biocides, and ion coatings have been the most commonly used chemical methods for biofilm prevention. They prevent biofilm formation by interfering with planktonic cell attachment and expansion of biofilm colonies on the surface of the device (26).

In the future, it will be very important while developing and designing any TMJR device that the compound or coating used to combat a biofilm is properly applied to the surface so that it does not wear off with function, and that the dosage of any anti-biofilm compound will be effective when delivered.

The characteristics of biocompatibility and bio-integration have made titanium alloy (Ti6AL4V) alloy the metal of choice for the major components of orthopedic and dental replacement devices. However, titanium alloys have low wear resistance and have been reported not to be appropriate as an articular bearing surface. Further, concern has been raised related to the release over time of corrosion wear related vanadium (V) and aluminum (Al) ions that have been reported associated with Ti6V4Al component instability in animal studies (27).

Titanium alloys are classified into the α, α+β, and β types based on the phase types existing in the alloy. β-phase titanium is the more ductile phase while the α-phase is stronger, but less ductile (28). A new class of titanium β-alloys may be promising candidates for the next generation of joint replacement implants. The advantages of newer titanium β-alloys are the incorporation of non-toxic alloying elements such as Nb, Mo, Ta, Zr and Sn in the formulation. These improved titanium alloys have demonstrated an elastic modulus closer to human bone, improved corrosion resistance and higher mechanical strength (29-31).

Despite its high strength and good resistance to corrosion, studies have reported that titanium wear particles and ions have been detected in local peri-implant tissue and remote organs (32). Titanium particles and ions are released from a component by mechanical wear, interaction bacteria biofilm, and inflammatory cells (33).

The disadvantages of low surface hardness, high friction coefficient, and unsatisfactory abrasive wear resistance of titanium alloy cannot be overlooked. These poor tribological properties raise a particular concern for biomedical load bearing implants, since they are in contact with corrosive bodily fluids and are subjected to functional movements which result in the release of metallic ions and wear debris into the body (34).

In order to overcome some of these disadvantageous properties of titanium, several surface modification techniques have been developed, such as organic and inorganic coatings, thermal oxidation, acid etching, electrochemical processes, plasma spraying and laser nitriding. Nano-scale texture appears to be more effective when compared with the macro- and micro- scale textures. Textured surface covered by coating/film has increasingly influenced and enhanced the performance of titanium alloy under functional loading (35).

Diamond-like carbon (DLC) film, a hard surface with diamond-like characteristics, is biocompatible and has high wear resistance, a low coefficient of friction, high chemical inertia, and bactericidal properties which can be enhanced using silver nanoparticles that provide effective reduction of bacterial adhesion in acrylic resin and interfaces. However, to date factors associated with DLC film deposition must be studied in order to obtain the best biocompatibility (36).

Alumina ceramic bearing materials attracted researchers due to their low friction, wettability, wear resistance, and biocompatibility. However, the first applications of alumina in orthopedics were associated with high fracture rates (37). Ceramic composite alternatives to zirconia, fabricated from mixtures of alumina and zirconia, zirconia-toughened alumina (ZTA), or alumina-toughened zirconia (ATZ) are suitable for orthopedic ceramic-on-polyethylene as well as ceramic-on-ceramic joint replacement devices. However, to date research regarding the in vivo performance of ZTA is still far from complete. Longer implantation studies are required to fully determine if ZTA components will outperform their counterparts in orthopedic devices (38). Whether DLC or alumina ceramic bearing surfaces are the future of TMJR remains to be determined based on laboratory and clinical testing outcomes.

Cobalt chrome (CoCr) alloy has proven over time to have excellent wear resistance and because of this property has been successfully utilized as the bearing surface for the mobile component in both orthopedic (femoral head) and TMJR (condylar head) devices when paired with UHMWPE. However, when paired with a CoCr alloy bearing surface (metal-on-metal), this combination has led to device failures due to excessive metal particle and ion related adverse local tissue reactions (13).

Cobalt chrome alloy as used in orthopedic and TMJR implants contains cobalt alloyed with 27–30% chromium, 5–7% molybdenum, with manganese and silicon <1%, iron <0.75%, nickel <0.5%, and with traces of carbon, nitrogen, tungsten, phosphorus, sulfur, and boron (39).

Dependent on their chemical composition, all metals implanted in the body undergo electrochemical related corrosion releasing free metal ions that can activate the immune system (40). Warshaw et al. reported that the most common allergen is nickel, affecting 15.5% of the general population (41). Foussereau and Laugier first reported the correlation between an eczematous dermatitis and orthopedic devices containing the CoCr alloy (42).

To date, there are few studies that report the incidence of TMJR metal hypersensitivity. Sidebottom and Mistry reported 39% of patients implanted with either TMJ, Inc (aka, Christensen, Golden, CO) CoCr alloy metal-on-metal or polymethylmethacrylate-on-metal TMJR devices demonstrated allergy to one or more of the metal components on skin patch testing (SPT) (43).

Hassan et al. questioned whether TMJR long-term outcomes are affected by the use of nonreactive metals in patients with documented preoperative metal hypersensitivity, and/or lead to the development of TMJR devices using new nonreactive materials (44).

De Meurechy and Mommaerts suggest exploration of the use of β-titanium alloys with better biocompatibility and wear properties, and an elastic modulus closer to bone, as well as alumina-toughened zirconia condyle bearing surfaces leading to the improved future TMJR devices (45).

Henry et al. recommended the use of an all titanium alloy condyle/fossa bearing to eliminate the use of nickel containing cobalt chrome alloy (46). The future of TMJR devices not only lies in the use of higher-quality CoCr, Ti, or Zr alloys as the embodiment of customized, patient fitted devices. The manufacture of TMJR systems may improve and become less costly as 3D printing techniques evolve (47).


Complications diagnosis and management in the future (Table 6)

Table 6

The most common complications associated with TMJR

1. Infection
2. Heterotopic ossification
3. Material hypersensitivity
4. Chronic post-TMJR pain

Mercuri LG. Complications Associated with TMJ TJR: Management and Prevention. In: Mercuri LG (ed). Temporomandibular Joint Total Joint Replacement – TMJ TJR – A Comprehensive Reference for Researchers, Material Scientists and Surgeons. New York: Springer, 2016;187-226. TMJR, temporomandibular joint replacement.

Infection

Despite the incidence of TMJR periprosthetic joint infection (PJI) being low (2.74%) (25), the clinical, psychological, and economic consequences are significant (48,49). Prevention of TMJR PJI includes reducing patients’ bacterial burden (50), administering prophylactic antibiotics one hour before surgery (51), and deterrence of biofilm formation on TMJR components by coating them with antimicrobial drugs or bactericidal nanocrystals (52,53).

While TMJR PJI prevention has proven to be the most effective strategy, making an appropriate well-timed diagnosis is not only critical, but can be the most challenging aspect of PJI management. A PJI is difficult to diagnose because to date there is no uniform or standardized criteria for the diagnosis. This is further complicated by the difficulty of differentiating a PJI from an adverse local tissue reaction, material hypersensitivity, aseptic joint failure, or a localized but unrelated skin reaction (54).

To date, there is no PJI diagnostic test that provides definitive organism sensitivity and specificity data. New molecular and genomic based assays are being developed in orthopedics. To determine their validity, these assays are being adapted to a diagnostic algorithm and subjected to clinical testing (55). However, many involve synovial fluid sampling which is a difficult undertaking in TMJR PJI cases. Future TMJR PJI research must investigate and determine the how to utilize these synovial fluid based molecular and genetic assays.

Heterotopic ossification (HO)

After infection, acquired HO is the next most common post-TMJR implantation complication (1.24%). Despite some evidence to support their use in the management of TMJR HO, NSAIDs, radiation therapy, and the autogenous fat graft have been considered as acceptable management and prophylactic options until further research elucidates improved alternatives (56).

Researchers have demonstrated a critical role for a naturally occurring glycoprotein, Alpha 2-Heremans-Schmid glycoprotein/fetuin-A (Ahsg fetuin-A) as an inhibitor of unwanted mineralization indicating further investigation may provide the use of fetuin-A as a novel therapeutic concept to prevent HO (57-60). Further research into this and other inhibitors of heterotopic bone and their delivery systems will undoubtedly follow.

Material hypersensitivity

Hypersensitivity reactions to implant materials, although uncommon, have been reported in the literature and, if symptomatic, require correct diagnosis and appropriate management (61). Wear and corrosion of functional joint implants introduce metal particles as well as metal ion debris into the adjacent tissues (62). Typically, a minor amount of metal debris does not prompt an adverse local tissue response. However, persons with a known hypersensitivity to a metal that is a component of an implant may respond differently. Contact with metal wear debris for them can result in immune cell activation, division, and the encouragement of the contribution of other immune cells in the reaction, thereby initiating an inflammatory response. Depending on the intensity of the person’s sensitivity to the material and the amount of wear debris generated, the response can be mild to severe (63,64). The resultant inflammatory response can also lead to damaging effects at the bone-implant interface resulting in symptoms of reflex muscle pain, skin eruptions and facial edema in TMJR patients (65). Chronic exposure to sensitizing metal debris can also result in protracted peri-implant inflammation which can lead to osteolysis around fixation screws, component micromotion and loosening, and catastrophic device failure (61,66).

The two most common diagnostic tests employed to determine material hypersensitivity are the SPT, and the lymphocyte transformation test (LTT). SPT is purely subjective and the testing process potentially can sensitize the patient to the metals being tested. Reports have demonstrated that false-negative SPT results may be due to the inability to reproduce on the skin the biological conditions to which a deep implant is exposed. Antigen responding cells in the skin (Langerhans cells) have different antigen-specific reactivity compared to those cells (T-lymphocytes) that respond to a deep implant (67,68). The LTT is an objective assay of material sensitivity based on the analysis of a patient’s T-lymphocytes to specific allergens in vitro elicited from a simple blood draw (69).

If, prior to TMJR a patient provides a history of intolerance to jewelry or an allergic reaction to a prior metal implant, LTT material hypersensitivity testing should be considered. However, to date, routine testing is not supported by the literature (40,70). Future investigation into the immune response to implanted materials leading to improved diagnosis and management and/or development of less reactive implant materials are required.

Post TMJR pain

The estimated mean incidence of chronic post-surgical pain has been reported between 10 and 50% for procedures such as thoracotomy, breast, inguinal hernia, and amputation surgery (71). However, it has rarely been reported post TMJR (0.43%) (72). However, when it occurs after TMJR, it is not only a problem for the patient, but also poses a diagnostic dilemma for the surgeon.

Acute pain typically derives from nociception in somatic or visceral tissues (intrinsic pain). However, not every pain sensation originates from an injury related stimulus (extrinsic pain) (73). There are both intrinsic and extrinsic that can cause pain after TMJR (figure 11). The surgeon must rule out one or the other in a systematic manner in order to manage the situation appropriately (48).


Intrinsic causes for post TMJR pain (Table 7)

Table 7

Intrinsic causes for post TMJR pain

1. Infection
2. Heterotopic ossification
3. Dislocation
4. Material hypersensitivity
5. Aseptic component or screw loosening
6. Component or screw fracture
7. Osteolysis
8. Neuroma formation
9. Synovial entrapment syndrome

Mercuri LG. Complications Associated with TMJ TJR: Management and Prevention. In: Mercuri LG (ed). Temporomandibular Joint Total Joint Replacement – TMJ TJR – A Comprehensive Reference for Researchers, Material Scientists and Surgeons. New York: Springer, 2016;187-226. TMJR, temporomandibular joint replacement.

The "Biologic Response to Metal Implants" document (74) states that when working up a patient with a painful total joint, hypersensitivity should be the last item on the differential diagnosis list, since the literature clearly demonstrates that 1% or less of joint replacement device failures are causally related to material hypersensitivity. Infection, heterotopic ossification, micromotion and loose hardware are much more common and should be ruled out first (75).

However, two other potential intrinsic causes have recently surfaced that may account for post-TMJR pain and therefore should be ruled out before TMJR removal and replacement—synovial impingement, and wear related, non- hypersensitivity local adverse tissue response.

Synovial impingement syndrome is a known cause of pain and dysfunction after orthopedic joint replacement (76-79). After joint replacement, a pseudosynovium develops. Westermark et al. and Monje et al. demonstrated this in TMJR patients (80,81). Further, Murakami et al. demonstrated synovial invaginations of synovial tissue in the TMJ similar to those found in the hip and knee (82) (See Murakami, et al. in this issue) If the pseudosynovium invaginates and becomes trapped between the bearing surface of a joint replacement device it can become inflamed and painful.

Davis et al. presented arthroscopic images of inflamed pseudosynovial tissue entrapped between the bearing surface of TMJR devices in patients with post TMJR pain and dysfunction. The results demonstrated that arthroscopic excision of the inflamed entrapped tissue led to decreased pain and increasing mandibular range of motion. However, the authors advised that future larger long-term studies were needed to further classify and define causes of post-TMJR chronic pain, the role of synovial entrapment in post-TMJR pain, and stressed that TMJR arthroscopy be reserved only for surgeons with Level 3 arthroscopy skills (83).

Although rare in TMJR due to minimal functional loading compared to hips and knees, material wear results in UHMWPE particulation and metal ion release leading to an adverse local tissue inflammatory response within the joint unrelated to material hypersensitivity (32,84). Therefore, this could be another source of post TMJR pain and dysfunction. Diagnostic local anesthesia infiltration may be diagnostic and careful debridement can be therapeutic. Further investigation into TMJR adverse local tissue response is certainly warranted.


Extrinsic causes for post TMJR Pain (Table 8)

Table 8

Extrinsic causes for post TMJR pain

1. Prior misdiagnosis
2. Chronic centrally mediated pain
3. Persistent myofascial/muscular pain
4. Complex regional pain syndrome I
5. Neurologic injury (CPRS II)
6. Temporalis tendonitis
7. Coronoid impingement
8. Frey’s neuralgia
9. Integrin formation

Mercuri LG. Complications Associated with TMJ TJR: Management and Prevention. In: Mercuri LG (ed). Temporomandibular Joint Total Joint Replacement – TMJ TJR – A Comprehensive Reference for Researchers, Material Scientists and Surgeons. New York: Springer, 2016; 187-226. TMJR, temporomandibular joint replacement.

It is noteworthy that many of these painful and dysfunctional post TMJR patients have been multiply operated and/or misdiagnosed muscular TMJ dysfunction patients with multiple comorbidities and persistent centrally mediated muscle pain.

A multicenter, cross-sectional study of the pre-operative risk factors associated with pain after total hip and knee replacement conveyed data that moderate to severe pain was reported by 20% at rest and 33% with activity. Predictors of post-implant pain at rest were female gender, severe pre-implant pain requiring the use of pre-implant opioids. Predictors of post-implant pain with activity were severity of the pain pre-implant, the use of anticonvulsants and anti-depressants, and prior hip/knee surgery (85).

A prospective analysis of data from total knee replacement patients identified potential predictors of outcomes. The most robust predictors were reported to be pre-implant pain/function, the less severe the pre-implant disease the better the outcome; diagnosis, rheumatoid arthritis patients did better than osteoarthritis patients; deprivation, patients from poorer areas had worse outcomes; and anxiety/depression, patients diagnosed with anxiety/depression had poorer outcomes (86).

The orthopedic literature also demonstrates that the greater the number of pre-implant co-morbidities, the poorer the outcomes (87-90). These results are harmonious with similar TMJ disorder data that showed that co-morbid conditions may explain why 50% of patients seeking care for TMJ pain, some of whom were multiply operated and/or exposed to failed materials or devices, still report experiencing pain 5 years later and 20% of chronic TMJ pain patients experience long-term disability from that pain (91-93).

The appropriate overall management of patients requiring TMJR requires the surgeon has made the right diagnosis, understands the patient associated predictors of outcomes and the comorbid conditions discussed above. The surgeon must then execute the surgical plan at the right time utilizing the appropriate TMJR device. This assures the results will be professionally satisfying for the surgeon, and most importantly will provide the best outcome for the patient. Improved understanding of the role of pre-TMJR co-morbid conditions on post-TMJR outcomes using patient reported data should be investigated in the future.


Clinical outcomes

Presently and certainly in the future, it should be required that all TMJR devices, whether custom or stock, wherever and however they are manufactured, undergo rigorous biomechanical laboratory and clinical testing, the results of which are reviewed by the appropriate regulatory body for safety and efficacy before marketing, sales, and use. Elledge et al. reported that as of the date of their publication, only 4 of the 27 devices presented had received regulatory agency approval. (10).

The 2 US TMJR devices, TMJ Concepts and Zimmer Biomet have demonstrated the kind of biomechanical and clinical outcomes testing, reporting and analysis by a government regulatory agency (US FDA) that should be mimicked in the future by all TMJR device manufacturers and the regulatory bodies to whom they report to assure the safety and efficacy of any prospective TMJR devices. Both of these devices have also continued to report successful clinical outcomes, as well as post-market surveillance studies demonstrating long-term safety and efficacy (94-97).

In the future, additive manufacturing (3D printing), augmented reality (AR), artificial intelligence (AI) and robot-assisted surgery (RAS) will become important tools in the growth and development of oral and maxillofacial surgery. What role these technologies will play in the future of TMJ disorder diagnosis, non-surgical and surgical management as well as TMJR device development and manufacturing remain to be determined.


Conclusions

Salvage management of end-stage TMJ pathologic conditions are considered indications for TMJR. The primary goal of TMJR is the long-term safe and effective restoration of mandibular function and form. In order to provide successful function and form long-term outcomes, any present or future TMJR device must be capable of managing the divergent functional, anatomic, aesthetic situations encountered. Therefore, all present and future TMJR devices must be able to demonstrate that they consist of biologically compatible materials, are designed and manufactured to bear the functional loads that will be delivered to their components, and that they are biomechanically and clinically analyzed and tested to assure their long-term safety and efficacy for patients.

“The arrogance of success is to think that what you did yesterday will be sufficient for tomorrow.”—C. William Pollard


Acknowledgments

Funding: None.


Footnote

Provenance and Peer Review: This article was commissioned by the Guest Editors (Stephen Feinberg and Louis Mercuri) for the series “Temporomandibular Joint Disorders Diagnosis and ManagementWhat Does the Future Hold?” published in Frontiers of Oral and Maxillofacial Medicine. The article was sent for external peer review organized by the Guest Editors and the editorial office.

Conflicts of Interest: The author has completed the ICMJE uniform disclosure form (available at https://fomm.amegroups.org/article/view/10.21037/fomm-2020-tjddm-02/coif). The series “Temporomandibular Joint Disorders Diagnosis and ManagementWhat Does the Future Hold?” was commissioned by the editorial office without any funding or sponsorship. LGM served as the unpaid Guest Editor of the series. LGM reports personal fees from TMJ Concepts, Ventura, CA USA, other from TMJ Concepts, Ventura, CA USA, during the conduct of the study; personal fees from MJ Concepts, Ventura, CA USA, other from MJ Concepts, Ventura, CA USA, outside the submitted work.

Ethical Statement: The author is accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.

Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0/.


References

  1. Mercuri LG. Temporomandibular Joint Replacement. In: Fonseca RJ, Carlson ER, Ness GM. editors. Oral and Maxillofacial Surgery. Philadelphia: Elsevier, 2018;923-40.
  2. Mercuri LG. End-Stage TMD and TMJ Reconstruction. In: Miloro M, Ghali G, Larsen P, et al. editors. Peterson’s Principles of Oral & Maxillofacial Surgery. 3rd Ed. Shelton, CT: PMPH, USA Ltd, 2012;1173-86.
  3. Onoriobe U, Miloro M, Sukotjo C, et al. How many temporomandibular joint total joint replacements will be performed in the United States by 2030? J Oral Maxillofac Surg 2016;74:1531-8. [Crossref] [PubMed]
  4. Kurtz S, Mowat F, Ong K, et al. Prevalence of primary and revision total hip and knee arthroplasty in the United States from 1990 through 2002. J Bone Joint Surg Am 2005;87:1487-97. [PubMed]
  5. Lotesto A, Miloro M, Mercuri LG, et al. Are oral and maxillofacial surgery residents trained adequately in alloplastic total temporomandibular joint replacement? J Oral Maxillofac Surg 2016;74:712-8. [Crossref] [PubMed]
  6. Momin M, Miloro M, Mercuri LG, et al. Survey of senior oral and maxillofacial surgery resident confidence in performing invasive temporomandibular joint procedures. J Oral Maxillofac Surg 2017;75:2091.e1-2091.e10. [Crossref] [PubMed]
  7. Lantada AD, Morgado PL. General Considerations for the Development of Biomedical Devices. In: Lantada AD (ed). Handbook on Advanced Design and Manufacturing Technologies for Biomedical Devices. New York: Springer, 2013;19-45.
  8. Katti KS. Biomaterials in total joint replacement. Colloids Surf B Biointerfaces 2004;39:133-42. [Crossref] [PubMed]
  9. Mercuri LG. The use of alloplastic prostheses for temporomandibular joint reconstruction. J Oral Maxillofac Surg 2000;58:70-5. [Crossref] [PubMed]
  10. Elledge R, Mercuri LG, Attard A, et al. Review of emerging temporomandibular joint total joint replacement systems. Br J Oral Maxillofac Surg 2019;57:722-8. [Crossref] [PubMed]
  11. Premarket ApprovalFDA.. (PMA). Available online: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P980052
  12. Premarket ApprovalFDA.. (PMA). Available online: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P020016
  13. Mercuri LG. editor. Temporomandibular Joint Total Joint Replacement - TMJ TJR - A Comprehensive Reference for Researchers, Material Scientists and Surgeons. New York: Springer, 2016.
  14. Kashi A, Chowdhury AR, Saha S. Finite element analysis of a TMJ implant. J Dent Res 2010;89:241-5. [Crossref] [PubMed]
  15. Hsu JT, Huang HL, Tsai MT, et al. Effect of screw fixation on temporomandibular joint condylar prosthesis. J Oral Maxillofac Surg 2011;69:1320-8. [Crossref] [PubMed]
  16. Ramos A, Relvas C, Completo A, et al. The importance of the geometry of the screws in TMJ implant fixation. Comput Methods Biomech Biomed Engin 2013;16:258-9. [Crossref] [PubMed]
  17. Tack P, Victor J, Gemmel P, et al. 3D-printing techniques in a medical setting: a systematic literature review. Biomed Eng Online 2016;15:115. [Crossref] [PubMed]
  18. Lethaus B, Bloebaum M, Koper D, et al. Interval cranioplasty with patient-specific implants and autogenous bone grafts—success and cost analysis. J Cranio-Maxillofac Surg 2014;42:948-51. [Crossref]
  19. Mashiko T, Otani K, Kawano R, et al. Development of three-dimensional hollow elastic model for cerebral aneurysm clipping simulation enabling rapid and low-cost prototyping. World Neurosurg 2015;83:351-61. [Crossref] [PubMed]
  20. 3D Printing Sucks: 5 Problems With 3D Printing and How to Fix Them. Available online: https://www.autodesk.com/redshift/5-problems-with-3d-printing-and-how-to-fix-them/
  21. AMFG. 10 of the Biggest Challenges in Scaling Additive Manufacturing for Production [Expert Roundup]. Available online: https://amfg.ai/2019/10/08/10-of-the-biggest-challenges-in-scaling-additive-manufacturing-for-production-expert-roundup/
  22. Mercuri LG. Microbial biofilms - a potential source of alloplastic device failure. J Oral Maxillofac Surg 2006;64:1303-9. [Crossref] [PubMed]
  23. Mercuri LG, Psutka D. Peri-operative, Post-operative and prophylactic use of antibiotics in alloplastic total temporomandibular joint replacement surgery: A survey and preliminary guidelines. J Oral Maxillofac Surg 2011;69:2106-11. [Crossref] [PubMed]
  24. Mercuri LG. Avoiding and managing temporomandibular joint total joint replacement surgical site infections. J Oral Maxillofac Surg 2012;70:2280-9. [Crossref] [PubMed]
  25. Mercuri LG. Prevention and detection of TMJ prosthetic joint infections update. Int J Oral Maxillofac Surg 2019;48:217-24. [Crossref] [PubMed]
  26. Roy R, Tiwari M, Donelli G, et al. Strategies for combating bacterial biofilms: A focus on anti-biofilm agents and their mechanisms of action. Virulence 2018;9:522-54. [Crossref] [PubMed]
  27. Geetha M, Singh AK, Asokamani R, et al. Ti based biomaterials, the ultimate choice for orthopaedic implants—a review. Prog Mater Sci 2009;54:397-425. [Crossref]
  28. Sankaran KK, Mishra RS. Titanium Alloys. In: Sankaran KK, Mishra RS. editors. Metallurgy and Design of Alloys with Hierarchical Microstructures. New York: Elsevier, 2017;21-40.
  29. Banerjee D, Williams J. Perspectives on titanium science and technology. Acta Mater 2013;61:844-79. [Crossref]
  30. Bahl S, Devadiga A, Chatterjee K, et al. Phase Stability Study of Recently Developed Metastable p‐Ti ALLOY Ti‐32Nb‐2Sn for Orthopedic Applications. In: Proceedings of the 13th World Conference on Titanium. Wiley Online Library, 2016;935-8.
  31. Bahl S, Krishnamurthy AS, Suwas S, et al. Controlled nanoscale precipitation to enhance the mechanical and biological performances of a metastable β Ti-Nb-Sn alloy for orthopedic applications. Mater Des 2017;126:226-37. [Crossref]
  32. Bijukumar DR, Souza JCM, Segu A, et al. Systemic and local toxicity of metal particles released from hip prostheses: A review of experimental approaches. Nanomedicine 2018;14:951-63. [Crossref] [PubMed]
  33. Reclaru L, Ardelean LC, Grecu AF, et al. Multicomponent Alloys for Biomedical Applications. IntechOpen. Available online: http://dx.doi.org/10.5772/intechopen.88250
  34. Lin N, Li D, Zou J, Xie R, et al. surface texture-based surface treatments on ti6al4v titanium alloys for tribological and biological applications: A mini review. Materials (Basel) 2018;11:487. [Crossref] [PubMed]
  35. Çahaa J, Alvesa AC, Affonço LJ, et al. Corrosion and tribocorrosion behaviour of titanium nitride thin films grown on titanium under different deposition times. Surf Coat Technol 2019;374:878-88. [Crossref]
  36. Silva AM, Figueiredo VMG, Prado RFD, et al. Diamond-like carbon films over reconstructive TMJ prosthetic materials: Effects in the cytotoxicity, chemical and mechanical properties. J Oral Biol Craniofac Res 2019;9:201-7. [Crossref] [PubMed]
  37. Clarke IC, Manaka M, Green DD, et al. Current status of zirconia used in total hip implants. J Bone Joint Surg Am 2003;85-A:73-84. [Crossref] [PubMed]
  38. Kurtz SM, Kocagöz S, Arnholt C, et al. Advances in zirconia toughened alumina biomaterials for total joint replacement. J Mech Behav Biomed Mater 2014;31:107-16. [Crossref] [PubMed]
  39. Standard Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Castings and Casting Alloy for Surgical Implants (UNS R30075). Available online: https://www.astm.org/Standards/F75
  40. Mercuri LG, Caicedo MS. Material hypersensitivity and alloplastic temporomandibular joint replacement. J Oral Maxillofac Surg 2019;77:1371-6. [Crossref] [PubMed]
  41. Warshaw EM, Maibach HI, Taylor JS, et al. North American contact dermatitis group patch test results: 2011-2012. Dermatitis 2015;26:49-59. [Crossref] [PubMed]
  42. Foussereau J, Laugier P. Allergic eczemas from metallic foreign bodies. Trans St Johns Hosp Dermatol Soc 1966;52:220-5. [PubMed]
  43. Sidebottom AJ, Mistry K. Prospective analysis of the incidence of metal allergy in patients listed for total replacement of the temporomandibular joint. Br J Oral Maxillofac Surg 2014;52:85-6. [Crossref] [PubMed]
  44. Hassan S, Mercuri LG, Miloro M. Does Metal Hypersensitivity Have Relevance in Patients Undergoing TMJ Prosthetic Replacement? J Oral Maxillofac Surg 2020;78:908-15. [Crossref] [PubMed]
  45. De Meurechy N, Mommaerts MY. Alloplastic temporomandibular joint replacement systems: a systematic review of their history. Int J Oral Maxillofac Surg 2018;47:743-54. [Crossref] [PubMed]
  46. Henry A, Inverso G, Granquist EJ. Revision temporomandibular joint arthroplasty for the treatment of acquired metal allergy and review of the literature. Int J Oral Maxillofac Surg 2020;49:356-60. [Crossref] [PubMed]
  47. He G, Eckert J, Dai QL, et al. Nanostructured Ti-based multi-component alloys with potential for biomedical applications. Biomaterials 2003;24:5115-20. [Crossref] [PubMed]
  48. Mercuri LG. Complications associated with TMJ TJR: management and prevention. In: Mercuri LG. editor. Temporomandibular joint total joint replacement—TMJ TJR— a comprehensive reference for researchers, material scientists and surgeons. New York. New York: Springer, 2016;187-200.
  49. Mercuri LG. Infection Following Total Joint Replacement. In: Bouloux G. Editor. Complications of TMJ Surgery. New York: Springer, 2017;135-47.
  50. Bryce E, Wong T, Forrester L, et al. Nasal photodisinfection and chlorhexidine wipes decrease surgical site infections: a historical control study and propensity analysis. J Hosp Infect 2014;88:89-95. [Crossref] [PubMed]
  51. Ban KA, Minei JP, Laronga C, et al. American College of Surgeons and Surgical Infection Society: surgical site infection guidelines, 2016 update. J Am Coll Surg 2017;224:59-74. [Crossref] [PubMed]
  52. Norambuena GA, Patel R, Karau M, et al. Antibacterial and biocompatible titanium-copper oxide coating may be a potential strategy to reduce periprosthetic infection: an in vitro study. Clin Orthop Relat Res 2017;475:722-32. [Crossref] [PubMed]
  53. Guan M, Chen Y, Wei Y, et al. Long-lasting bactericidal activity through selective physical puncture and controlled ions release of polydopamine and silver nanoparticle loaded TiO(2) nanorods in vitro and in vivo. Int J Nanomedicine 2019;14:2903-14. [Crossref] [PubMed]
  54. Deirmengian C, Kardos K, Kilmartin P, et al. Diagnosing periprosthetic joint infection: has the era of the biomarker arrived? Clin Orthop Relat Res 2014;472:3254-62. [Crossref] [PubMed]
  55. Mercuri LG. Temporomandibular joint replacement periprosthetic joint infections: a review of early diagnostic testing options. Int J Oral Maxillofac Surg 2014;43:1236-42. [Crossref] [PubMed]
  56. Mercuri LG, Saltzman BM. acquired heterotopic ossification in alloplastic joint replacement. Int J Oral Maxillofac Surg 2017;46:1562-68. [Crossref] [PubMed]
  57. Schafer C, Heiss A, Schwarz A, et al. The serum protein Alpha 2-Heremans-Schmid glycoprotein/fetuin-A is a systemically acting inhibitor of ectopic calcification. J Clin Invest 2003;112:357-66. [Crossref] [PubMed]
  58. Rittenberg B, Partridge E, Baker G, et al. Regulation of BMP-induced ectopic bone formation by Ahsg. J Orthop Res 2005;23:653-62. [Crossref] [PubMed]
  59. Albilia JB, Tenenbaum HC, Clokie CML, et al. Serum levels of BMP-2, 4, 7 and AHSG in patients with degenerative joint disease requiring total arthroplasty of the hip and temporomandibular joints. J Orthop Res 2013;31:44-52. [Crossref] [PubMed]
  60. Fuery MA, Liang L, Kaplan FS, et al. Vascular ossification: Pathology, mechanisms, and clinical implications. Bone 2018;109:28-34. [Crossref] [PubMed]
  61. Hallab N. Material Hypersensitivity. In: Mercuri LG. editor. Temporomandibular joint total joint replacement—TMJ TJR— a comprehensive reference for researchers, material scientists and surgeons. New York: Springer, 2016;227-50.
  62. Mathew MT, Abbey S, Hallab NJ, et al. Influence of pH on the tribocorrosion behavior of CpTi in the oral environment: synergistic interactions of wear and corrosion. J Biomed Mater Res B Appl Biomater 2012;100:1662-71. [Crossref] [PubMed]
  63. Caicedo MS, Pennekamp PH, McAllister K, et al. Soluble ions more than particulate cobalt-alloy implant debris induce monocyte costimulatory molecule expression and release of proinflammatory cytokines critical to metal-induced lymphocyte reactivity. J Biomed Mater Res A 2010;93:1312-21. [PubMed]
  64. Wang Y, Dai S. Structural basis of metal hypersensitivity. Immunol Res 2013;55:83-90. [Crossref] [PubMed]
  65. Wolford LM. Factors to consider in joint prosthesis systems. Proc (Bayl Univ Med Cent) 2006;19:232-8. [Crossref] [PubMed]
  66. Jacobs JJ, Hallab NJ. Loosening and osteolysis associated with metal-on-metal bearings: A local effect of metal hypersensitivity? J Bone Joint Surg Am 2006;88:1171-2. [PubMed]
  67. Hallab N, Merritt K, Jacobs JJ. Metal sensitivity in patients with orthopaedic implants. J Bone Joint Surg Am 2001;83:428-36. [Crossref] [PubMed]
  68. Luque I, Leyva L, Jose Torres M, et al. In vitro T-cell responses to beta-lactam drugs in immediate and nonimmediate allergic reactions. Allergy 2001;56:611-8. [Crossref] [PubMed]
  69. Hallab NJ. Lymphocyte transformation testing for quantifying metal-implant-related hypersensitivity responses. Dermatitis 2004;15:82-90. [Crossref] [PubMed]
  70. Jacobs JJ. Clinical manifestations of metal allergy. Adverse reactions to byproducts of joint replacements (AAOS/ORSI). Presented at: Annual Meeting of the American Academy of Orthopaedic Surgery. San Francisco, CA, February 7-11, 2012.
  71. Schnabel A, Pogatzki-Zahn E. Predictors of chronic pain following surgery. What do we know? Schmerz 2010;24:517-31. [Crossref] [PubMed]
  72. TMJ Concepts, Ventura CA, personal communication, 2017.
  73. Merskey H, Bogduk N. Classification of chronic pain. Descriptions of chronic pain syndromes and definitions of pain terms. 2nd ed. Seattle: IASP, 1994.
  74. www.fda.gov. Biological Responses to Metal Implants. U.S. Food and Drug Administration. Center for Devices and Radiological Health.
  75. Mercuri LG. Complications Associated with TMJ TJR: Management and Prevention. In: Mercuri LG. editor. Temporomandibular Joint Total Joint Replacement - TMJ TJR - A Comprehensive Reference for Researchers, Material Scientists and Surgeons. New York: Springer, 2016;187-226.
  76. Dupont JY. Synovial plicae of the knee. Controversies and review. Clin Sports Med 1997;16:87-122. [Crossref] [PubMed]
  77. Kent M, Khanduja V. Synovial plicae around the knee. Knee 2010;17:97-102. [Crossref] [PubMed]
  78. Funk L, Levy O, Even T, et al. Subacromial plica as a cause of impingement in the shoulder. J Shoulder Elbow Surg 2006;15:697-700. [Crossref] [PubMed]
  79. Ruiz de Luzuriaga BC, Helms CA, Kosinski AS, et al. Elbow MR imaging findings in patients with synovial fringe syndrome. Skeletal Radiol 2013;42:675-80. [Crossref] [PubMed]
  80. Westermark A, Leiggener C, Aagaard E, et al. Histological findings in soft tissues around temporomandibular joint prostheses after up to eight years of function. Int J Oral Maxillofac Surg 2011;40:18-25. [Crossref] [PubMed]
  81. Monje F, Mercuri LG, Villanueva-Alcojol L, et al. Synovial metaplasia found in tissue encapsulating a silicone spacer during 2-staged temporomandibular joint replacement for ankylosis. J Oral Maxillofac Surg 2012;70:2290-8. [Crossref] [PubMed]
  82. Murakami K, Hori S, Yamaguchi Y, et al. Synovial plicae and temporomandibular joint disorders: surgical findings. J Oral Maxillofac Surg 2015;73:827-33. [Crossref] [PubMed]
  83. Davis CM, Hakim M, Choi DD, et al. Early clinical outcomes of arthroscopic management of the failing alloplastic temporomandibular joint prosthesis. J Oral Maxillofac Surg 2020;78:903-7. [Crossref] [PubMed]
  84. Mercuri LG, Urban RM, Hall DH, Mathew MT. Adverse local tissue responses to temporomandibular joint implants. J Oral Maxillofac Surg 2017;75:2076-84. [Crossref] [PubMed]
  85. Liu SS, Buvanendran A, Rahtmell JP, et al. Predictors for moderate to severe acute postoperative pain after total hip and knee replacement. Int Orthop 2012;36:2261-7. [Crossref] [PubMed]
  86. Judge A, Arden NK, Cooper C, et al. Predictors of outcomes of total knee replacement surgery. Rheumatology (Oxford) 2012;51:1804-13. [Crossref] [PubMed]
  87. Jones CA, Voaklander DC, Suarez-Almazor ME. Determinants of function after total knee arthroplasty. Phys Ther 2003;83:696-706. [Crossref] [PubMed]
  88. Lingard EA, Katz JN, Wright EA. Predicting outcome of total knee arthroplasty. J Bone Joint Surg Am 2004;86:2179-86. [Crossref] [PubMed]
  89. Escobar A, Quintana JM, Bilbao A. Effect of patient characteristics on reported outcomes after total knee replacement. Rheumatology (Oxford) 2007;46:112-9. [Crossref] [PubMed]
  90. Arendt-Nielsen L, Nie H, Laursen MB. Sensitization in patients with painful knee osteoarthritis. Pain 2010;149:573-81. [Crossref] [PubMed]
  91. Drangsholt M, LeResche L. Temporomandibular disorder pain. In: Crombie IK, Croft PR, Linton SJ, et al. editors. Epidemiology of pain: a report of the task force on epidemiology of the international association for the study of pain. International Association for the Study of Pain. Seattle, 1999;203-33.
  92. Velly AM, Look JO, Carlson C, et al. The effect of catastrophizing and depression on chronic pain - a prospective cohort study of temporomandibular muscle and joint pain disorders. Pain 2011;152:2377-83. [Crossref] [PubMed]
  93. Lim PF, Maixner W, Kahn AA. Temporomandibular disorder and comorbid conditions. JADA 2011;142:1365-7. [Crossref] [PubMed]
  94. Wolford LM, Mercuri LG, Schneiderman ED, et al. Twenty-year follow-up study on a patient-fitted temporomandibular joint prosthesis: the Techmedica/TMJ Concepts device. J Oral Maxillofac Surg 2015;73:952-60. [Crossref] [PubMed]
  95. FDA. 522 Postmarket Surveillance Studies. Available online: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfPMA/pss.cfm?t_id=40&c_id=393